ACU20FD High Pressure

The ACU20FD digital mass flow meter consists of a mass flow sensor, a laminar flow element, a flow control valve, and an amplification and control circuit. It uses digital signal input and output and features a short preheating time, small zero drift, and high reliability.

Range: 0-100 SLM

Category: Gas Flow

Overview

Applications

Specifications

Dimensions

Software

◆Thermal principle, higher accuracy, faster response, with network port

◆Tubular flow path, not easy to block

◆Suitable for various low-pressure and high-pressure pipelines

◆Short preheating time, small zero drift, high reliability

◆Analog input/output can be used to control and display flow rate with a secondary instrument or flow computer

ACU20FD High-Precision Mass Flow Meter/Flow Controller It adopts the principle of capillary heat transfer method to measure the mass flow rate of gas, which is not affected by temperature and pressure. The flow signal measured by the sensor is amplified and then compared with the set voltage. The obtained difference is used to drive the control regulating valve, and the closed-loop control of the flow rate through the channel makes it equal to the set flow rate.

Glass industrySolar energyAtmospheric monitoringIndustrial manufacturingPetroleum and petrochemical
Coal and metallurgyGas production and distributionEnvironmental protectionVarious instrument analysis
SeriesACU20FD
ModelLCLMMCMM
Measurement Range0-(2,5,10...1000) SCCM, 0-(1,5,10...30) SLM0-(30,40,50...300) SLM
Accuracy±0.5%F.S
Linearity±0.5%F.S
Repeatability±0.2%F.S
Response Time<0.2s<0.1s<0.2s<0.1s
Operating Differential Pressure0.1~0.5MPa (specified when ordering)
Measurement Range Ratio50:1100:150:1100:1
Analog Signal0~5V、4~20mA
Digital SignalRS232/485,MODBUS protocol
Power Supply24VDC
Communication ProtocolModbus、DeviceNet、profinet、EtherCAT、profibus, etc.
Operating Temperature0~50℃
Maximum Withstand Pressure40MPa/70MPa (specified when ordering)
Electrical ConnectionDB9 port
DisplayTouch screen
Leakage Rate1×10⁻⁹Pa m³/S
Protection LevelIP40
ValveValve normally closed/Valve normally closed/
Base MaterialStainless steel
Sealing MaterialViton, nitrile rubber, EPDM, perfluoroether
Body ThreadG3/8 inner thread
ConnectorCompression fitting, pagoda, quick-connect, VCR, flange, etc.

ACU20FD-LC40MPa Dimensions

ACU20FD-LM (40Mpa high pressure) Dimensions

Mass flow measurement and control system V5.0